An automated external defibrillator (AED) works by analyzing the heart's electrical signal through electrode pads, deciding whether a shockable rhythm is present, and then delivering a controlled burst of electrical current that simultaneously depolarizes heart muscle cells so the heart's natural pacemaker can re-establish a coordinated rhythm. The device charges a capacitor to a preset energy level, typically 120–200 joules, and discharges it across the chest in under 10 milliseconds. Everything else — the voice prompts, the pad placement diagrams, the legal protections — exists to make that one biophysical event happen correctly and fast.

The chain is: pads sense voltage across the chest → filters isolate the ECG signal → an algorithm classifies the rhythm → the capacitor charges → a biphasic waveform crosses the myocardium → cells reset → a perfusing rhythm either returns or it does not. Each link has failure modes, and understanding them changes how you use the device.

The sensing pathway: what the pads actually measure

Defibrillation pads are not microphones for electricity. They are large-area electrodes, roughly 8–12 cm across, with conductive gel that lowers skin impedance from thousands of ohms to roughly 50–150 ohms. Skin is a poor conductor; the gel and the pad's surface area convert a high-resistance barrier into a low-resistance window.

The AED measures the voltage difference between the two pads. Because the pads sit on opposite sides of the heart — one typically below the right clavicle, the other on the left lateral chest — the vector between them passes through the ventricles. That geometry is why pad position matters more than pad brand.

Inside the device, an analog front end amplifies the millivolt-scale cardiac signal and applies filters. A high-pass filter removes baseline wander from chest compressions and breathing. A low-pass filter strips muscle noise. What remains is a cleaned electrocardiogram, sampled digitally and fed to the rhythm-analysis algorithm. The device is, in effect, a single-lead ECG monitor with a decision tree attached.

The algorithm: how the AED decides to shock

The core question the algorithm answers is binary: is this rhythm shockable or not? Only two rhythms are treated with a shock — ventricular fibrillation (VF) and pulseless ventricular tachycardia (VT). Both represent disorganized or excessively fast electrical activity in the ventricles with no effective pumping.

The algorithm does not diagnose. It computes features: rate, amplitude, regularity, wave morphology, and the proportion of the signal within certain frequency bands. VF is chaotic, with varying amplitude and no consistent QRS complexes. Asystole is nearly flat. Normal sinus rhythm has regular, repeating complexes. The device compares these computed features against thresholds derived from annotated rhythm databases, and it is deliberately biased toward specificity — it would rather withhold a shock from a shockable rhythm than shock a perfusing one.

That bias explains the device's most counterintuitive behavior. An AED can refuse to shock a heart that is dying in fine VF, because low-amplitude VF can resemble asystole to a filtering algorithm. Conversely, CPR artifact, pad loosening, or patient movement can mimic a shockable rhythm and prompt an inappropriate charge. Modern devices mitigate this with motion sensors and continuous artifact detection, but the failure mode exists.

A professional caveat worth internalizing: the algorithm analyzes a single lead over a short window. It has no access to pulse, blood pressure, or patient responsiveness. It cannot distinguish pulseless VT from VT with a pulse. That distinction is your job, and it is why the device prompts you to check for normal breathing and signs of life before and after analysis.

The shock itself: energy, waveform, and why biphasic won

When the algorithm classifies a shockable rhythm, the device charges its capacitor. Charging takes roughly 5–15 seconds depending on the model and energy setting. The capacitor stores energy in an electric field; discharging it through the patient releases that energy as current over a very short interval.

The critical variable is not energy in joules but current through the myocardium, which depends on transthoracic impedance. A 150-joule shock delivered across 50 ohms produces far more current than the same shock across 150 ohms. Modern AEDs measure impedance through the pads before discharge and adjust the delivered waveform to compensate.

Waveform shape matters as much as magnitude. Older monophasic devices delivered a single current pulse in one direction. Biphasic devices deliver current in one direction, then reverse it partway through the pulse. Biphasic waveforms achieve comparable or better defibrillation success at lower energy, which reduces myocardial damage. This is why most current AEDs default to 150–200 joules biphasic rather than the 200–360 joules typical of monophasic units.

Feature Monophasic AED Biphasic AED
Waveform Single direction Reverses mid-pulse
Typical energy 200–360 J 120–200 J
Myocardial injury at equivalent success Higher Lower
First-shock success (VF, historical ranges) Moderate Moderate-to-high
Current availability Largely legacy Standard
Pad compatibility Device-specific Device-specific
Impedance compensation Limited Built into waveform

The table's numbers are ranges because actual delivered energy varies by model and by measured impedance. Do not assume a higher number means a stronger shock to the patient; the device is managing current, not impressing you with joules.

The physiology: why one shock can restart a heart

Ventricular fibrillation is a circuit problem. Instead of contracting together, millions of ventricular muscle cells fire in chaotic loops. No coordinated squeeze occurs, so no blood moves. The heart is electrically active and mechanically dead.

Defibrillation works by overwhelming that chaos. The shock depolarizes a critical mass of myocardial cells simultaneously, driving them into a refractory state where they cannot conduct. For a brief window — milliseconds to a few seconds — the chaotic loops are extinguished. If enough cells are reset, the heart's own pacemakers, the sinoatrial node and the conduction system, can resume control and generate an organized rhythm.

Two variables govern success. The first is the mass of myocardium depolarized. Too few cells reset, and the fibrillation loops re-establish. The second is the state of the myocardium itself. An ischemic, acidotic, or energy-depleted heart is harder to defibrillate and more likely to re-enter VF. This is why CPR before and after the shock matters: compressions keep the coronary arteries perfused, which improves the odds that a reset heart stays reset.

Plain-language version: the shock does not "jump-start" the heart like a car battery. It silences a chaotic electrical argument so the heart's own conductor can take over again.

The time course: minutes that decide outcomes

The following table reflects the well-established relationship between time to defibrillation and survival in out-of-hospital cardiac arrest with a shockable rhythm. Survival figures are approximate ranges drawn from the resuscitation literature and vary by response system, bystander CPR rate, and patient factors. Treat them as directional, not as guarantees.

Elapsed time from collapse Typical rhythm state Approximate survival with CPR + AED What is happening physiologically
0–1 min VF, high amplitude Highest Myocardium still oxygenated; shock often succeeds
1–3 min VF, high amplitude High Oxygen reserve falling; CPR preserves perfusion
3–5 min VF transitioning Moderate Amplitude declining; acid load rising
5–10 min Coarse to fine VF Lower Myocardium energy-depleted; shock less effective
10–15 min Fine VF or asystole Low Algorithm may refuse to shock fine VF
>15 min Asystole common Very low Resuscitation rarely succeeds without rapid advanced care

The clinical takeaway: every minute of delay degrades both the heart's shockability and the patient's chance of meaningful survival. The AED's value is not just the shock; it is the speed at which a non-clinician can deliver one.

Where the device stops: limitations and false signals

An AED is a narrow tool. It treats exactly two rhythms and does nothing for the rest.

Cardiac arrest caused by pulseless electrical activity (PEA) or asystole will not receive a shock, correctly. These rhythms require CPR, cause correction, and advanced interventions. An AED that stays silent on a patient in PEA is functioning as designed.

The device also cannot fix the underlying cause. If the arrest stems from a massive hemorrhage, a pulmonary embolism, a tension pneumothorax, or severe hypoxia, defibrillation may restore a rhythm that immediately fails again. The shock is necessary but not sufficient.

Practical warnings worth memorizing:

Do not delay compressions to retrieve or attach an AED if another rescuer is present; send someone else. Do not touch the patient during analysis or shock delivery — motion corrupts the rhythm read and creates a shock hazard. Do not place pads over a transdermal medication patch, an implanted pacemaker generator, or a wet chest without drying it; each can alter current delivery. Do not use an AED on a patient who is responsive, breathing normally, or has a palpable pulse. Do not assume a "no shock advised" message means the patient is fine; it means the rhythm is not shockable and CPR should continue. Do not stop CPR to watch the device; resume immediately after any shock or no-shock decision.

Practical actions: what to do, in order, with thresholds

The following sequence reflects current lay-rescuer guidance from major resuscitation bodies. It is not patient-specific medical advice.

Confirm unresponsiveness and abnormal or absent breathing. Gasping is not normal breathing and does not rule out cardiac arrest. Call emergency services or direct a specific bystander to call, and send another to fetch the AED. Give the dispatcher the address and a callback number. Start chest compressions immediately at 100–120 per minute, 5–6 cm deep, allowing full recoil. If trained, add rescue breaths at a 30:2 ratio. Turn on the AED as soon as it arrives and follow its voice prompts. Expose and dry the chest, remove medication patches, and shave heavy chest hair only if it prevents pad adhesion. Attach pads per the diagrams — typically right upper chest and left lateral chest — and clear everyone during analysis. Deliver the shock if advised, then resume compressions immediately without checking for a pulse. The device will re-analyze on its own schedule. Continue until EMS takes over, the patient shows signs of life, or you are physically unable to continue.

Thresholds that matter: compressions at 100–120 per minute, depth of at least 5 cm but not more than 6 cm in adults, full recoil between compressions, and minimal interruptions of under 10 seconds. These are consensus targets, not absolutes, and they apply to adult out-of-hospital arrest.

What the evidence does and does not support

AEDs improve survival in out-of-hospital cardiac arrest when used early and combined with CPR. That conclusion is robust across public-access defibrillation programs and is the basis for widespread placement in airports, gyms, schools, and workplaces. The mechanism is well characterized: rapid defibrillation of VF before it degrades to asystole.

What the evidence does not support is the idea that an AED replaces CPR or that a shock alone saves lives. Survival depends on the interaction of time, rhythm, myocardial condition, and bystander action. Devices vary in energy protocols and algorithm tuning, and those differences matter at the margins, but no consumer-facing AED removes the need for prompt compressions.

If you want to be useful in a cardiac arrest, the highest-yield preparation is not memorizing device specs. It is knowing where the nearest AED is, how to start compressions without hesitation, and how to direct other people to help. The device handles the electricity. You handle the minutes.